· Ceramic sensor with diaphragm Al2O3, 99.9%
(Standard)
Ceramic thick sensor, Silicon sensor (Optional) · Corrosive or abrasive applications · Short reaction time · Extremely high overload limit · Range: -0.1~ 70MPa · Option: LCD display
Technical data:
* Supply voltage :12...45 VDC
* 2-wire-system :4...20 mA with superimposed
signal for HART protocol digital communication
* 3-wire-system (optional):1...5V
* Signal range:3.8mA.....22.8mA
* Signal on alarm:3.8mA/ 22.8mA/Others on request
* Accuracy :Nominal range 0.1%
0.25% accuracy for range trundown 4:1
0.5% accuracy for range trundown 10:1
* Power supply effect:Negligible
* Vibration effect : <0.01% of URL /g when tested 200Hz in any axis relative
* Thermal effec :±0.05%/10K
* Compensated temperature range of sensor:-20...80℃
* Permissible load:≤50(V-12)Ω
* Stability:0.1% FSO/Year
* Switch on delay :5s
* Damping :0 to 100s, step: 0.1s
* Response time:200 ms (without consideration of electronic damping)
* Self stability configuration :0 to 2%
* Filter configured:0 to 160 uA
* Communication resistance for HART Protocol:Typ. 250 Ω
Areas of application:
*Absolute pressure and gauge pressure in gases,
Steams or liquids in all areas of process engineering
and process measurement technology.